Volume 21, number 1
 Views: (Visited 444 times, 1 visits today)    PDF Downloads: 180

Bagad K. S, Bacchao K, Bagade S. B, Amrutkar R. D, Patil D. D. RP-HPLC Estimation of Clobetasol Propionate and Salicylic Acid using Quality by Design Approach. Biotech Res Asia 2024;21(1).
Manuscript received on : 28-11-2023
Manuscript accepted on : 19-02-2024
Published online on:  13-03-2024

Plagiarism Check: Yes

Reviewed by: Dr. Nicolás Padilla-Raygoza

Second Review by: Dr. Mohammed Oday Ezzat

Final Approval by: Dr. Ghulam Md Ashraf

How to Cite    |   Publication History    |   PlumX Article Matrix

RP-HPLC Estimation of Clobetasol Propionate and Salicylic Acid using Quality by Design Approach

Kunal S. Bagad1, Kunal Bacchao1, Shashikant B. Bagade2, Rakesh D. Amrutkar3 and Dipak D. Patil3*

1Pharmaceutical Quality Assurance Department, H. R. Patel Institute of Pharmaceutical Education and Research, Shirpur. Dist. Dhule. Maharashtra, India.

2Pharmaceutical Chemistry, SVKM's NMIMS, School of Pharmacy and Technology Management, Mumbai-Agra Highway No. 3, Sawalde, Babulde, Post. Gidhade,Tq.-Shirpur. Dist- Dhule, Maharashtra India.

3Pharmaceutical Chemistry Department, K. K. Wagh College of Pharmacy, Nashik. Hirabai Haridas Vidyanagari, Amrutdham, Panchavati, Nashik. Maharashtra, India.

Corresponding Author E-mail: dipakpatil888@gmail.com

DOI : http://dx.doi.org/10.13005/bbra/3225

ABSTRACT: The RP-HPLC method for CLOP and SA estimation from bulk and pharmaceutical dosage from has been developed and validated. For analytical methods to be robust, current ICH guidelines, Q8 to Q11 were suggested use of analytical quality by design (AQbD) includes adoption of current systematic approaches. The proposed method was optimized and developed using Taguchi orthogonal design. The RP-HPLC method parameters were optimized by box- Behnken design. The stationary phase used C18 Princeton column (150mm × 4.6mm × 5µm) with acetonitrile: 0.05M phosphate buffer (pH 2.5, adjustment with by ortho - phosphoric acid) as mobile phase at ratio of 60:40v/v, 1.0 ml/min of flow rate along with UV-Visible wavelength of detection 240 nm. The linearity over concentration 5-15 µg/ml for CLOP and 600-1500 µg/ml for SA (r2 =0.9969 for CLOP and 0.9943 for SA) was found. The retention time for SA was 2.2 min. and CLOP 7.0 minute. The % recovery was found to be 98.0.3 SA and 97.84 or CLOP. As per ICH analytical method validation guidelines [Q2 (R1)], the RP-HPLC method was validated.

KEYWORDS: Clobetasol propionate, Ointment analysis, RP-HPLC Salicylic acid

Download this article as: 
Copy the following to cite this article:

Bagad K. S, Bacchao K, Bagade S. B, Amrutkar R. D, Patil D. D. RP-HPLC Estimation of Clobetasol Propionate and Salicylic Acid using Quality by Design Approach. Biotech Res Asia 2024;21(1).

Copy the following to cite this URL:

Bagad K. S, Bacchao K, Bagade S. B, Amrutkar R. D, Patil D. D. RP-HPLC Estimation of Clobetasol Propionate and Salicylic Acid using Quality by Design Approach. Biotech Res Asia 2024;21(1). Available from: https://bit.ly/3Vgj14x

Introduction

Clobetasol propionate is clobetasol’s 17-O-propionate ester. It’s a powerful corticosteroid that’s used in treatment of eczema and psoriasis, among other skin conditions. It works as an anti-inflammatory agent. Clobetasol propionate exerts its action by binding to cytoplasmic glucocorticoid receptors and then stimulates glucocorticoid receptor mediated expression of genes (Figure 1). This causes the stopping of inflammatory mediator’s production to reduce while anti-inflammatory proteins production to rise. CLOP produces phospholipase A2 inhibitory proteins which causes stoppage of anti-inflammatory precursors release like arachidonic acid from membrane phospholipids 1-3.

Figure 1: Clobetasol propionate

Click here to view Figure

Salicylic acid, also known as 2-hydroxy benzoic acid, is an antibacterial, antifungal, and keratolytic agent. Warts, psoriasis, corns, and other skin problems are treated with it. It softens and loosens dry, scaly, or thickened skin, allowing it to slip off or be readily removed.

Salicylic acid permanently blocks the activity of COX-1 and COX-2, which results in reduced production of prostaglandins and thromboxanes from arachidonic acid. Because of its analgesic and anti-inflammatory properties, salicylate is used to treat rheumatic disorders due to their analgesic and anti-inflammatory properties4,5.

Figure 2: Salicylic acid

Click here to view Figure

As per the Literature, many analytical methods like RP-HPLC, UV-Visible Spectrophotometry, Colorimetry, GC, etc. had found for the SAL and CLO determination in combination with many other drugs2,6-11. The RP-HPLC methods reported earlier for this combination were gradient, longer Rt for CLO as 18.70 min.12 and require column temperature 45°C with short Rt for SAL as 2.25 min2. Also, there is no analytical method with analytical quality by design approach has been reported for SAL and CLO combination. The 6% SAL and 0.0.5 % CLO formulation (Manufactured by Besto chem. Ltd. Delhi) marketed by Liva Healthcare Ltd. available to treat eczema, psoriasis with other skin condition issues.

Material and methods                              

Chemicals and reagents

Clobetasol propionate a gift sample obtained from Orbicular Pharmaceutical Ltd, Hyderabad. Salicylic Acid a laboratory sample obtained from research-lab fine chem industries, Mumbai.

The (HPLC grade) solvents used were methanol, acetonitrile, water whereas chemicals were of analytical grade like sodium dihydrogen phosphate monohydrate (Merck), and sodium hydroxide (Research fine lab).

Instrumentation

Chromatographic analysis was carried out on Agilent (1200) Series) with EZ Chrome Elite 3.3.2 software. The stationary phase was C18 (Princeton) (4.6 mm x 150 mm x 5 mm ),  acetonitrile: Phosphate buffer (pH 2.5) as mobile phase with detection at 240 nm. The DOE software used was Design-Expert 13.

Methods

Preparation of Mobile phase and physical standard stock solutions

It was found that acetonitrile, and sodium dihydrogen phosphate buffer (pH 2.5) satisfactory resolution as compared to other mobile phases. The optimum composition of mobile phase utilized was acetonitrile, 0.05 M sodium dihydrogen phosphate buffer (pH 2.5) (60:40) with flow rate 1ml/min, wavelength of detection 240 nm at room temperature. The mobile phase was filtered to remove particulate matter and sonicated to degas before its use. The standard stock solution of SA and CLOP were prepared using methanol. The mix working physical standard solution containing 10 μg/ml CLOP and 100 μg/ml of SA was prepared with appropriate dilution.

Linearity study

The aliquots solution of (0.05-0.15ml) of standard stock solution of clop and (0.6-1.5 ml) of SA was diluted to 10 ml of mobile phase. The last concentration in range of (5-15μg/ml) for CLOP and (600-1500 μg/ml) for SA solution was obtained by dilution with acetonitrile. Each concentration was analyzed in triplicate.

Pharmaceutical formulation analysis

The 2 gm of ointment has been weighed that contain (120 mg of SA and 1 mg of CLOP) and add 30 ml n-hexane solution and add 5 ml methanol two times and shake continuously 15-20 min. and separate the two layers in separator. After lower layer withdrawn, sonicated and diluted to 25.0 ml with mobile phase. Then 2.5 ml withdrawn diluted with mobile phase to 10.0 ml which is equivalent to (1200 μg/ml) SA and (10 μg/ml) CLOP.

Results and Discussion

Wavelength selection

The appropriate selection of wavelength of detection in HPLC-UV analysis improves sensitivity of method. The wavelength at which both drugs show good response is called an ideal wavelength for the analysis drugs. The SA and CLOP standard solutions scanned over range of 200-400 nm. For simultaneous determination of both drugs from ointment dosage form, the wavelength 240 nm was used because both drugs had appropriate absorbance at 240 nm. (Figure 3).

Figure 3: Overlay spectrum of UV spectrum of SA and CLOP

Click here to view Figure

Optimization of chromatographic parameters by using experimental design

Screening by Taguchi orthogonal model

Taguchi orthogonal model was used to screen the effect of chromatographic parameters organic content of mobile phase (A), salt of concentration (B), pH of aqueous phase (C), flow usrate (D), column type (E), solvent type (F), pH modifier (G). On the basis of initial experiments, the range of value used in the design was A: 40-60 v/v, B:10-50, C:2.5-5.5pH, D: 0.8-1.2ml/min, E: C8-C18 F: acetonitrile and methanol F:10-50mM as shown in Table 1. The experiments were run and responses were number of theoretical plates, retention time, and tailing factor.

Table 1: Experimental factors for Taguchi orthogonal model

Experimental factors Code level low High
Organic content of mobile phase (%) A 40 60
salt of concentration B 10 50
pH of aqueous phase C 2.5 5.5
 flow rate D 0.8 1.2
column type E C8 C18
solvent type F Methanol Acetonitrile
pH modifier G Sodium Potassium

 The experimental general in Taguchi orthogonal model

The method variables were screened by Taguchi orthogonal design as shown in Table 2. The main goal of this study was to pinpoint the key factors that have a significant impact on method performance using a smaller number of experiments.

Table 2: Design matrix of Taguchi orthogonal model

Run Organic content of mobile phase (%) salt of concentration pH of aqueous phase flow rate column type solvent type pH modifier
1 40 50 5.5 0.8 C8 Acetonitrile Na+ salt
2 60 10 5.5 0.8 C18 Methanol Na+ salt
3 40 10 2.5 1.2 C18 Acetonitrile Na+ salt
4 60 50 2.5 0.8 C18 Acetonitrile K+ salt
5 60 50 2.5 1.2 C8 Methanol Na+ salt
6 40 50 5.5 1.2 C18 Methanol K+ salt
7 60 10 5.5 1.2 C8 Acetonitrile K+ salt
8 40 10 2.5 0.8 C8 Methanol K+ salt

This design was used to evaluate the retention time, tailing factor, resolution, no. theoretical plates responses as shown in Table 3.

Table 3: Responses Taguchi orthogonal model

 

Run

Response 1

Retention time

Response 2

Tailing factor

Response 3

Resolution

Response 4

No. of theoretical plates

  SA CLOP SA CLOP SA CLOP SA CLOP
1 2.30 14.3 1.45 2.27 0 3.90 2569 93
2 5.64 50 1.79 5 0 30 2436 2000
3 3.5 20.4 1.57 1.23 0 31.75 2280 11465
4 0.9 4.0 0.90 1.43 0 13.22 479 3262
5 5.6 50 1.29 5 0 30 1527 2000
6 1.68 50 1.73 5 0 30 1177 2000
7 2.12 6.98 1.20 1.19 0 15.81 814 7722
8 2.32 12.60 1.68 1.33 0 15.25 1394 2102

 The method variables were screened by Taguchi orthogonal design. The half normal and Pareto chart results shows solvent type, flow rate, pH and organic phase composition in mobile phase expressively affects critical quality attributes. In solvent acetonitrile, retention time of SA was 1.7-2.5 min. as compared to methanol. The overall performance of all design only affects the solvent factor.

Box-Behnken design

The critical method variables which affects critical method attributes were optimized using Box-Behnken design. A three level box- Behnken design with five center replication was applied for optimization of the organic phase content of mobile phase (A), pH of aqueous phase (B), flow rate (C) as the Table No. 4 show parameters and responses13-14.

Table 4: Design matrix of BBD model for optimizing chromatographic variables.

Run

No.

Factor 1 Factor 2 Factor 3  

                                  Responses

  organic phase content (%) pH flow rate Retention time Tailing factor resolution No.of theoretical plates
SA CLOP SA CLOP  CLOP SA CLOP
1 50 4 1.2 1.4 1.88 1.07 1.88 12.52 2177 1943
2 50 5.5 1 1.6 5.4 2.91 1.65 0 191 2355
3 60 4 1 1.6 4.5 2.1 1.5 0 214 2237
4 70 4 1.2 1.2 2.4 2.5 1.5 7.89 2572 2535
5 60 5.5 1.2 1.16 3.1 1.2 1.5 9.32 721 2655
6 60 5.5 0.8 1.7 4.4 1.15 1.41 9.52 847 2984
7 60 2.5 0.8 2 5.2 1.13 1.5 0 2151 2944
8 60 4 1 1.6 4.5 2.1 1.5 0 214 2237
9 70 5.5 1 1.5 2.7 2.15 1.51 6 1027 2827
10 60 2.5 1.2 1.4 2.9 2.02 1.58 5.68 329 2661
11 60 4 1 1.6 4.4 2.13 1.71 8.15 397 2414
12 50 4 0.8 30 10 5 1.89 0 300 1998
13 60 4 1 1.6 4.4 2.13 1.71 8.15 397 2414
14 60 4 1 1.6 4.4 2.13 1.71 8.15 397 2414
15 50 2.5 1 1.8 7.1 2.5 1.7 10.71 485 1836
16 70 4 0.8 1.9 3.7 1.97 1.5 8.12 1861 3076
17 70 2.5 1 1.7 3.1 0.93 1.43 5.04 399 3068

We used statistical multiple linear regression analysis to optimize the analytical data. The ANOVA and F test were used to select the best fitting second-order quadratic polynomial equation. In Table 5, the value of regression coefficient and their associated p- values are shown.

Table 5: Analysis of variance for response

Source Sum of square Df Mean square F – value P – value  
Model 53.87 6 8.98 15.56 0.0002 Significant
A-Organic phase content 19.47 1 19.47 35.75 0.0002  
B -pH 0.9112 1 0.912 1.58 0.2374  
C- flow rate 21.19 1 21.19 36.73 0.0001  
AB 0.42 1 0.42 0.73 0.4122  
AC 11.63 1 11.63 20.16 0.0012  
BC 0.2500 1 0.25 0.43 0.5252  
Residual 5.77 10 0.57  
Lack of fit 5.76 6 0.95 319.84 < 0.0001 Significant
Pure error 0.0120 4 0.0030  
Cor total 59.64 16  

It was observed that pH of aqueous phase significantly affects the response (p- value <0.005).

The main effect of organic content, pH of aqueous phase and flow rate

From the plots observed, the 60 % organic content, pH 2.5, flow rate 1 ml/min was investigated to obtain maximum responses.

Interaction plot of responses

The main effect plot for each factor was different for response.  The % of organic content was increases and pH of aqueous phase 2.5 till the responses i.e. retention time of salicylic acid to be increase as shown in Figure 4. The % of organic content was increases and pH of aqueous phase 2.5 till the responses i.e. tailing factor of salicylic acid to be increase. Figure 5.

Figure 4: Surface plot of RT vs organic content with aqueous phase pH.

Click here to view Figure

Figure 5: Surface plot of Tf vs organic content with aqueous phase pH.   

                             

Click here to view Figure

The % of organic content was 60% increases, aqueous phase pH 2.5 till the responses i.e. tailing factor of salicylic acid to be increase as shown in Figure 6. The % of organic content was 60% increases & pH of aqueous phase 2.5 till the responses i.e. retention time clobetasol propionate to be increase (4-5min.) as shown in Figure 7.

Figure 6: Surface plot of TP vs content with aqueous phase pH     

  

Click here to view Figure

Figure 7: Surface plot of RT vs organic organic content with aqueous phase pH

                  

Click here to view Figure

The % of organic content was 60% increases with aqueous phase pH 2.5 till the responses i.e. resolution for CLOP to be increase as shown in Figure 8. The % of organic content was 60% increases and pH of aqueous phase 2.5 till the responses i.e. tailing factor clobetasol propionate to be less than 2.0% as shown in Figure 9.

Figure 8: Surface plot of RS vs organic content with aqueous phase pH.

Click here to view Figure

Figure 9: Surface plot of Tf vs organic content with aqueous phase pH     

                          

Click here to view Figure

The % of organic content was 60% increases and pH of aqueous phase 2.5 till the responses i.e. theoretical plates clobetasol propionate to more than 2300 as shown in Figure 10.

Figure 10: Surface plot of RT vs organic content, and aqueous phase pH

Click here to view Figure

The effect of variables like A, B and C on Rt was shown with the help of 2D response surface plots. 

Optimized method parameters for experimentation

From the surface plot and interaction plots method parameters which makes the rugged method were chosen. The method to be considered rugged with respect to method parameters: organic content of mobile phase (A) (50-70) %, pH of aqueous phase (B) (2.5 – 5.5) and flow rate (C) (0.8 -1.0 ml/minutes.) with the resolution more than 3, tailing factor less than 2.0% and retention time to be acceptable within 8 minutes as design space.

Experimental design was used to optimize the specific method variables and effect of variables on the method attributes. The Table 6 has been shown the chromatographic condition and selection of stationary phase, mobile phase and all the parameters which was very critical to the method development for clobetasol propionate and salicylic acid estimation by using Box- Behnken design.

Table 6: optimized chromatographic condition by Box-Behnken design

Parameters Chromatographic condition
Stationary phase C18 Princeton (150 mm x 4.6mm x 5 µm)
Mobile phase Acetonitrile, phosphate buffer pH 2.5 (60:40)
Wavelength of analysis 240 nm
Flow rate 1 ml/min
Loop capacity 20 µl

By using optimized chromatographic condition from above Table No.6, the sample analysis of 10 μg/ml of CLOP and (10 μg/ml) of SA (10 μg/ml) was performed. The chromatogram was obtained is shown in below Figure 11.

Figure 11: HPLC Chromatogram of standard SA and CLOP

Click here to view Figure

Method validation

As per ICH Analytical method validation guidelines [Q2 (R1)], the proposed method was validated for parameters like accuracy, precision, linearity and range, LOD, LOQ, robustness, specificity etc. (ICH guidelines 2005).

System suitability

Using optimized chromatographic conditions, mix working standard solutions were analyzed. Each sample was analyzed five times. Table 7 displays the results of system suitability.

Table 7: Results of system of suitability

Sr.

No

Peak area Asymmetry Retention time RS Theoretical plates
  SA CLOP SA CLOP SA CLOP CLOP SA CLOP
1 128995 931198 1.94 1.51 1.74 4.45 9.33 1745 2345
2 123385 944235 1.93 1.54 1.7 4.43 9.19 1756 2384
3 130121 945224 1.85 1.56 1.69 4.28 8.96 1845 2436
4 128552 962541 1.86 1.49 1.69 4.23 9.23 1823 2311
5 130020 965414 1.89 1.52 1.74 4.33 9.45 1745 2345
Mean 128214.6 949722.4 1.894 1.524 1.712 4.344 9.232 1896 2364.2
SD 2487.639 12681.04 0.036111 0.024166 0.023152 0.084758 0.163021 42.56947 42.69614
%RSD 1.940215 1.335236 1.906597 1.585702 1.352317 1.951162 1.76583 2.245225 1.805944

Linearity study

The calibration curve was created by graphing the drug concentration against the peak area of SA and CLOP, respectively. The calibration curves are shown in Figure 12 for SA and Figure 13 for CLOP.

Figure 12: Salicylic acid calibration curve.

Click here to view Figure

Figure 13: Clobetasol propionate calibration curve

Click here to view Figure

The SA and CLOP were found linear with concentration range of 600-1500μg/ml with r2 value 0.9969 and concentration range of 5-15μg/ml with r2 value 0.9943 respectively.

Application of proposed method to laboratory mixture

Laboratory mixture analysis was conducted to see the feasibility of the optimized method for quantitative analysis. The % purity was found 98.41 ± 1.21 for SA with % RSD value 1.23 whereas 99.88 ± 1.40 for CLOP with % RSD value 1.43.

Application of proposed method to ointment analysis

The SA and CLOP from pharmaceutical ointment dosage form were analyzed by developed analytical method. The ointment formulation analysis results are depicted in the following Table 8.

Table 8: Results of ointment analysis

Drug Amount taken (μg/ml) % Amount found ±SD % RSD
SA 1200 99.56 ± 0.36 0.36
CLOP 10 95.78 ± 0.56 0.61

 Accuracy (% recovery)

The accuracy of analytical method for CLOP and SA estimation was determined at 80 %, 100 % and 120 % of the label claim. The accuracy of analytical method was determined by calculating the % recovery. The peak area was recorded of each analysis. The % recovery was found 98.03 ± 0.77 for SA with % RSD 0.78 and 97.84 ± 0.56 for CLOP with % RSD 0.58 respectively. The results shows the method to be accurate one reflecting with low % RSD values.

Precision, Repeatability study

The repeatability and intermediate precision study was used to determine precision study of method. The intermediate precision and repeatability study was executed by analyzing physical mixture. The inter-day precision % RSD for SA was 0.65 whereas 0.89 for CLOP. The intra-day precision % RSD for SA was 1.05 whereas 0.69 for CLOP. In repeatability study, % RSD values were 0.83 for SA whereas 0.48 for CLOP. The % RSD value less than two signifies the method is precise.

Specificity and Selectivity

The developed method was found to be selective and specific as no other component eluting and interfering at SA and CLOP retention time. There was no any interference from formulation components and both drugs and resolved properly. The base line also did not shown any significant noise. No any co-elution observed during analysis while studying with photo-diode array detector. The photodiode array detector was used to assess and confirm the peak purity of SA and CLOP. The summary of validation parameters for HPLC method as shown in Table 9. 

Table 9: Summary of validation parameters for HPLC method

Parameters Observations
SA CLOP
Linearity range (µg/ml) 600-1500 5-15
Regression equation Y = 201.27x +75896 Y = 13037x + 37238
Correlation coefficient 0.9969 0.9943
% recovery ± SD 98.03 ± 0.77 97.84 ± 0.56
LOD 40.79 3.20
LOQ 123.59 9.72
Intra-day precision

(% RSD)

0.65 0.89
Inter-day precision

(% RSD)

1.05 0.69
Specificity Specific Specific
Robustness Robust Robust

 Conclusion

The proposed RP-HPLC method was simple, accurate, reproducible and robust for estimation of salicylic acid (SA) and (CLOP) clobetasol propionate in bulk as well as ointment dosage form. The SA and CLOP % recovery values were found 98.29 and 98.46 % signifies accuracy of the developed method. The method showed high precision as evidenced by the low relative standard deviations at each level of the recovery experiment. The proposed method was validated, developed and optimized as per ICH Q2 (R1) guidelines and using Taguchi orthogonal design respectively. Box- Behnken design was used to optimize HPLC method parameters. Clobetasol Propionate and Salicylic Acid analysis from pharmaceutical dosage forms, the developed and optimized method is effective and is a valuable quality control tool.

Acknowledgement

The authors are grateful to the Principal, H. R. Patel Institute of Pharmaceutical Education and Research, Shirpur for providing necessary facilities to carry out the research work.

Conflict of Interest

The authors declare that there is no conflict of interest.

Funding Sources

NIL

Authors’ Contribution

K. S. Bagad, K. Bacchao, Dipak D. Patil were involved in study conception and design, data collection, S. B. Bagade, Dipak D. Patil in analysis and interpretation of results. R. D. Amrutkar, Dipak D. Patil in manuscript preparation along with all authors where all equally contributed.

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author.

Ethics Approval Statement

No any studies on human or animal was conducted hence it is not applicable.

References

  1. Esposito MC, Santos ALA, Bonfilio R, de Araújo MB. A critical review of analytical methods in pharmaceutical matrices for determination of corticosteroids. Critical reviews in analytical chemistry. 2020;50(2):111-24. https://doi.org/10.1080/10408347.2019.1581050
    CrossRef
  2. Patel B, Raj H, Jain V, Sutariya V, Bhatt M. Development and validation of reversed phase—High performance liquid chromatography method for clobetasol propionate and salicylic acid in its pharmaceutical dosage forms. Pharma Sci Monit. 2014;5(Suppl 1):374-85. https://www.pharmasm.com/current_issue2.php?archive=42
    CrossRef
  3. Patel N, Patel P, Meshram D. A Development and Validation of RP-HPLC Method For Simultaneous Estimation of Nadifloxacin and Clobetasol Propionate In Its Pharmaceutical Dosage Form. Am J PharmTech Res. 2016;6 (5) 21. https://ajptr.com/archive/volume-6/october-2016-issue-5
  4. Abass AM, Rzaij JM, ghalib Salman H, Al-Hashemi WKH. A Review on a Some Analytical Methods for Determination of Salicylic Acid. Open Access Journal of Chemistry, 3 ( 3), 2019, 22-28. https://doi.org/10.22259/2637-5834.0303005
    CrossRef
  5. Lévêque J, Corcuff P, Gonnord G, Montastier C, Renault B, Bazin R, et al. Mechanism of action of a lipophilic derivative of salicylic acid on normal skin. Skin Research and Technology. 1995;1(3):115-22. https://doi.org/10.1111/j.1600-0846.1995.tb00030.x
    CrossRef
  6. Ahmed NR, Mohamad NS. Spectrophotometric determination of clobetasol propionate in pharmaceutical preparations and environmental samples. World Journal of Pharmacy and Pharmaceutical Sciences. 2018;7(10):167-73. https://journals.indexcopernicus.com/api/ file/viewByFileId/695580.pdf
  7. Barange H, Asghar S, Gour P. Development of Analytical Method and Validation for Simultaneous Estimation of Clobetasol Propionate and Ketoconazole in Pharmaceutical Cream Formulation by RP-HPLC Method. INDO AMERICAN JOURNAL OF PHARMACEUTICAL SCIENCES. 2017;4(10). http://doi.org/10.5281/zenodo.1039704
  8. Devi N, Kumar S, Rajan S, Gegoria J, Mahant S, Rao R. Development and validation of UV spectrophotometric method for quantitative estimation of clobetasol 17-propionate. Asian Journal of Chemistry and Pharmaceutical Sciences. 2016;1(1):36-40. https://www.i-scholar.in/index.php/AJPAC/article/view/122258
    CrossRef
  9. Jakasaniya MA, Shah JS, Maheswari DG. Simultaneous estimation of clobetasol propionate and fusidic acid in cream dosage form by reversed phase high performance liquid chromatographic method. Pharmacophore. 2014;5(2):231-8. https://pharmacophorejournal.com/article/simultaneous-estimation-of-clobetasol-propionate-and-fusidic-acid-in-cream-dosage-form-by-reversed-phase-high-performance-liquid-chromatographic-method
  10. Kumaravel S, Raj D, Anbazhagan S, Shanmugapandiyan P. Simultaneous determination of halobetasol propionate and salicylic acid related sub-stances in ointment formulation and identifi-cation of impurities. International Journal of Pharmaceutics and Drug Analysis. 2016:276-80. https://www.ijpda.com/index.php/journal/article/view/235
  11. Manoharan G. Development and validation of a stability-indicating RP-HPLC method for the estimation of clobetasol propionate in bulk and ointment dosage form. Eur J Pharm Med Res. 2017;4:119-26. https://www.ejpmr.com/home/abstract_id/2342
  12. Bhuyian MHU, Rashid M, Islam A, Tareque M. Development and validation of method for determination of clobetasol propionate and salicylic acid from pharmaceutical dosage form by HPLC. British journal of Pharmaceutical Research. 2015;7:375-85. 10.9734/BJPR/2015/18494
    CrossRef
  13. Bonde S, Bonde C, Prabhakar B. Quality by design based development and validation of HPLC method for simultaneous estimation of paclitaxel and vinorelbine tartrate in dual drug loaded liposomes. Microchemical Journal. 2019;149:103982. https://doi.org/10.1016/j.microc.2019.103982
    CrossRef
  14. Patil KD, Bagade S, Bonde S. QbD-enabled stability-indicating assay method for the estimation of linezolid in newly developed gelatin nanoparticles for anti-tubercular therapy. Chromatographia. 2020;83(8):963-73. https://doi.org/10.1007/s10337-020-03925-9
    CrossRef
(Visited 444 times, 1 visits today)

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.